This section controls various configuration settings for A/D, P/R and Hardware acceleration.
Acquisition
- Try to use Fast Downloading
- If checked, the data download from the A/D will start immediately when the board is armed. If the data download rate is low, this may cause DMA timeout errors.
- Disable Burst DMA
- This option is set at the factory and should not be changed.
- Discard Follower Data
- If checked the preliminary TOF image will be discarded.
- Disable Advanced Follower features
- If checked, Follower gates will collect TOF data only.
- Match Thickness Type on Parent
- If checked, the Thickness type being collected will be matched on the parent Follower (e.g. Peak-to-Peak, Edge-to-Edge, etc), rather than always using the Edge crossing. This will be disabled if the "Disable Advanced Follower Features" option is checked.
- Ignore Peak Thresholds
- If checked peak amplitude data will be collected regardless of threshold size. Rising/Falling edge TOF is not valid while this checkbox is checked.
- Enable Gate Gains
- If checked, the user can adjust the gain on the waveform within a gate region.
- Always use Software Trigger when not scanning
- If checked, the digital oscilloscopes will always be triggered by ODIS when not in a scan, at a rate of approximately 20 times a second (this will vary depending on scope and gate settings). The trigger setting for the board is ignored until the scan starts. Whether or not a board actually triggers depends on if it is enabled and if appropriate trigger connections and pulser/receiver settings are present.
- Allow different trigger modes per channel
- If checked, the default trigger mode can be overridden by selecting the trigger modes via the Digital Oscilloscope.
- Don't trigger on channels that are not visible
- If checked, only the currently visible Scope tab will update when idle. Other channels will not be triggered.
- A/D Encoder resolutions
- This sets the counts per physical unit that the A/D board(s) use to to calculate increments when using Encoder Triggering.
P/R
- If Idle, Disable P/Rs
- This option can be used to turn off the P/R(s) after a certain time has passed without any activity detected. This helps prevent long-term wear on the pulser and transducers. Set to "Never" to keep the pulser on at all times. If a scan is running, this parameter is ignored.
- Clamp Pulser Energy
- This is a factory setting and should not be changed.