Scanning

From OKOS Wiki
Jump to navigation Jump to search

Scanning Settings

Scanning

Configures basic Scanning settings

Feature Example ODIS-IC? ODIS-NDT? Description
Axis N/A x x
Scanning Configuration Tab
Close Previous Collection on Scan Start
if checked will close the previous scan and ask if you’d like to save the data before closing. To close without being asked to save make sure Discard Previous Data is checked. If not checked all previous scans will be left open and accessible by the thumbnails on the right side of the main window.
Update Thumbnail during Scanning
allows you to see scan progression on the thumbnails in real time. This feature is useful if you have only one monitor or are using the second monitor for analysis of other images.
OverScan
Sets an amount of extra motion added to the scan length. This amount is added to both ends of Linear and Tray Scans. It ensures that the scanner will move beyond the encoder trigger region on each scan line.
Display A-Scans during Scan
Will update the oscilloscope during the scan. This is useful for real time correction of Follower and Gate placement/length
No. of Points Per Download
Number of points to collect on the A/D board before downloading to memory. This also sets how many triggers it takes for the scope to update during a scan. If the number of points per scan line is less than this amount, the scan will attempt to download at least once per line.
Changing Acceleration with Velocity
Change Acceleration when Velocity Changes
Connects the slider bar with both velocity and acceleration by a factor of your choosing.
Disable Updates during a Scan
if checked ODIS will not update the image of the current scan until after the scan has completed. If not checked will allow you to choose how many lines should be scanned before your image is updated. Disabling updates or increasing # of scan lines before update may speed up scan times on scans requiring larger data collection.
Align to Zero (Rotational Scans)
If checked, the system will attempt to find the Zero position on the turntable before scanning, and use said position to align the resulting image data. Note: this will only work on a turntable that is wired to send a "Zero" indication to the motor controller.
  • Software Only
If checked, the system will attempt to calculate the zero position of the Rotational axis to align the resulting image data.

See Also:Motor Trigger Connections,Triggering

After Scan, Return To...
Allows you to choose the place the transducer will return to after completing a scan.
  • Center – returns to the center of the scan area
  • Corner – returns to the top left corner of the scan area
  • Home – Returns to your system’s home position
  • None – leaves the transducer where it finishes.


Passive (ODIS-NDT onl)

Configures basic Scanning settings

Feature Example ODIS-IC? ODIS-NDT? Description
Axis N/A x x

Focusing

Feature Example ODIS-IC? ODIS-NDT? Description
Axis N/A x x
Focusing Configuration Tab
Reverse Gate Tracking Direction
When tracking is turned on, changes the direction the gates move when you raise or lower your Z-Axis.
Move Scope View when Gate Tracking
if checked moves the waveform when the focus changes. If not checked, leaves the wave form in place when changing focus.
Keep Focus Changes
will keep your current focus if changed during a scan.


Auto-Center

Allows you to choose the velocity of the auto-center procedure.

Feature Example ODIS-IC? ODIS-NDT? Description
Axis N/A x x

Tray Scans

Tray Scan Configuration Tab
Feature Example ODIS-IC? ODIS-NDT? Description
Axis N/A x x
Update Tray Scans Every Row
If checked, the Tray Scan image will only update after a row has completed. Otherwise, the Tray Scan will update every scan line (unless image updates are disabled entirely).
Wait for Row Step to Complete
If checked, the Scan Axis will not move until the jump between Tray rows has completed. Otherwise, scanning of the next row starts as soon as the row jump starts.
When Scanning Each Chip Individually
Water Path Calibration Range (System Units)
The distance the Focus Axis will move when calibrating the water path. If the transducer will be very close to the surface of the part, make sure this value is small.
Number of Water Path Tests
The number of checks that will be made to see if the follower is crossed when checking for Water Path.
Recall Water Path Timeout(seconds)
The amount of time the system will wait in order to get the Water Path values for Autofocus. If no results can be collected in the specified time period, Autofocus will not change the Focus axis position.


Warnings and Errors

Warnings and Errors Configuration Tab

Allows the operator to turn off or adjust pop-up warnings. These should be left on until users are comfortable with the software.

Feature Example ODIS-IC? ODIS-NDT? Description
Axis N/A x x


Show Warnings before Starting Scan
Uncheck to hide all warnings. Keep checked to specify which warnings will be displayed before the scan starts.
Scan Axis Warning
This will display the current scan axis before starting the scan. The user has an opportunity to cancel the scan if the wrong scan axis is set (e.g., Y-Axis instead of X-Axis)
Scan Start Change Warning
This will display a warning if the current position of the axes is not at a scan start or center. The user has the opportunity to set the current position as the scan start or center.
Focus Change Warning
This will display a warning if the current Focus axis is not at the position that was defined as the focus for the previous scan. The user will have the opportunity to recall the focus or keep the current one.
Range Warning
this will display a warning if the current scan is outside the range as specified in Motors.xml. The user will have an opportunity to cancel the scan.
Error Bounds
the maximum error allowed before a position error is triggered.
Scanline Timeout
when scanning, if the scan line collection takes more than the specified number of seconds, a timeout error will be generated.
On Scanline Error
the options here determine the reaction of the system when errors occur duing a scan.
Abort ScanStop the scan immediately and display an error message.
Ignore and ContinueIgnore the error and proceed with the next scan line regardless.
Rescan LineForces a rescan of the current line. There are two sub-options:
  • Reconcile: force a reconcile of the motor and encoder (based on current Reconcile options)
  • Lower Speed: reduce the scan velocity by the specified percentage.
Maximum Reconcile Error


Advanced Settings (Scanning)

Advanced Scan Settings Configuration Tab
Feature Example ODIS-IC? ODIS-NDT? Description
Axis N/A x x
Reconcile before Scan
If checked, the system will reconcile the motor and encoder positions before a scan. The way it reconciles depends on the Reconcile Options (see below).
Check End of Motion
If checked, the system will wait until the step axis has come to a complete stop before starting the next scan line.
End of Motion Delay
This forces the system to wait the specified number of milliseconds before starting the next scan line.
Use Quick Setup
If checked, The Simple Mode interface will have additional checkboxes added that enable and disable A/D boards, as well as set them for External triggering. Should only be used with a system that has multiple A/D boards.
Enable Surface Mapping
If checked, allows for the mapping of a warped image for contour scanning. Note: drift may occur and repeat scans may have varying results if contour scan kit is not present on your system.
"On Reconcile" options
The user can reconcile in two ways: set the Motor Position to the current Encoder position, or vice-versa. It is recommended that the user set the Motor Position to the Encoder position unless otherwise directed by OKOS.